Chest Radiograph Stand


Safe and stable
Chest X-ray frame series
  •  X1

    1. Up and down movement range of the cassette along the column:

    400mm-1850mm, with a deviation of not more than ±5%, and mechanical limit.

    2. Size dimensions of the cassette: 530x494x75mm

    3. Able to clamp various models of flat panel detectors and cassette dimensions.

    4. The grid can be installed on the side (498.5x449mm).

    5. Overall height of the chest radiograph stand column: 2060mm.

    6. Up and down movement handbrake, and hand-release automatic brake.

  • X2

    1. Movement: 350mm-1850mm.

    2. Operation mode: manual, and electromagnetic brake.

    3. Cassette dimensions: 17×17 detector.

    4. With AEC ionization chamber assembly position and grid insertion position.

    5. Able to detect the presence or absence of grids: yes.

    6. Able to detect the presence or absence of detectors: yes.

    7. Dimensions: 380×530×2200Mm (length × width × height).

    8. Equipped with a potentiometer for measuring up and down movement.

  • X3

    1. Movement: 350mm-1850mm

    2. Operation mode: manual and electric integration, at 60mm/s

    3. Cassette dimensions: 17×17 detector

    4. With AEC ionization chamber assembly position and grid insertion position

    5. Power-off protection function: yes

    6. Able to detect the presence or absence of grids: yes

    7. Able to detect the presence or absence of detectors: yes

    8. Dimensions: 380×530×2200mm (length × width × height)

    9. Equipped with a potentiometer for measuring up and down movement

  • X4

    1. Movement: 350mm-1850mm

    2. Operation mode: manual and electric integration, at 60mm/s

    3. Cassette dimensions: 17×17 detector

    4. With AEC ionization chamber assembly position and grid insertion position

    5. Power-off protection function: yes

    6. Able to detect the presence or absence of grids: yes

    7. Able to detect the presence or absence of detectors: yes

    8. Detector rotation angle: -20°~90°

    9. Dimensions: 760×530×2200mm (length × width × height)

    10. Equipped with a potentiometer for measuring up and down movement

  • X5

    1. Movement: 350mm-1850mm
    2. Operation mode: manual and electric integration, at 60mm/s
    3. Cassette dimensions: 17×17 detector
    4. With AEC ionization chamber assembly position and grid insertion position
    5. Power-off protection function: yes
    6. Able to detect the presence or absence of grids: yes
    7. Able to detect the presence or absence of detectors: yes
    8. Detector rotation angle: -20°~90°
    9. Dimensions: 640×530×2260mm (length × width × height)
    10.Equipped with a potentiometer for measuring up and down movement



[parameter]


1. Braking force: ≥100N

2. Starting force: ≥100N

3. The moving range of the cassette

along the column: 400- 1850 mm。


[Storage conditions and service life of the device]

1. Ambient temperature: -20℃ ~ 55 ℃;

2. Relative humidity: 10% ~ 93%,

including condensation;

3. Atmospheric pressure: 50kPa ~ 106kPa;









Shenzhen Angell Technology Co., Ltd.

Room 201, Building G3, TCL International E City, Shuguang Community, Xili Street, Nanshan District, Shenzhen, China

  • Tel: (+86) 755-86966744
  • Fax: (+86) 755-26859389
  • Email: market@szangell.com